The MAST motional Stark effect diagnostic

Loading...
Thumbnail Image

Date

Authors

Conway, N. J.
De Bock, M. F. M.
Michael, C. A.
Walsh, M. J.
Carolan, P. G.
Hawkes, N. C.
Rachlew, E.
McCone, J. F. G.
Shibaev, S.
Wearing, G.

Journal Title

Journal ISSN

Volume Title

Publisher

American Institute of Physics (AIP)

Abstract

A motional Stark effect (MSE) diagnostic is now installed and operating routinely on the MAST spherical tokamak, with 35 radial channels, spatial resolution of ∼2.5 cm, and time resolution of ∼1 ms at angular noise levels of ∼0.5°. Conventional (albeit very narrow) interference filters isolate π or σ polarized emission. Avalanche photodiode detectors with digital phase-sensitive detection measure the harmonics of a pair of photoelastic modulators operating at 20 and 23 kHz, and thus the polarization state. The π component is observed to be significantly stronger than σ, in reasonably good agreement with atomic physics calculations, and as a result, almost all channels are now operated on π. Trials with a wide filter that admits the entire Stark pattern (relying on the net polarization of the emission) have demonstrated performance almost as good as the conventional channels. MSE-constrained equilibrium reconstructions can readily be produced between pulses.