Cultural advice

The Australian National University acknowledges, celebrates and pays our respects to the Ngunnawal and Ngambri people of the Canberra region and to all First Nations Australians on whose traditional lands we meet and work, and whose cultures are among the oldest continuing cultures in human history.

Aboriginal and Torres Strait Islander peoples are advised that ANU Library collections may include images, names, voices, and other representations of deceased persons.

Material in the collection may contain terms, language or views that reflect the period in which the item was created and may be considered inappropriate today.

Tunneled Hemodialysis Catheter Tip Design and Risk of Catheter Dysfunction: An Australian Nationwide Cohort Study

Loading...
Thumbnail Image

Date

Authors

Lazarus, Benjamin
Polkinghorne, Kevan
Gallagher, Martin
Coggan, Sarah
Gray, Nicholas
Talaulikar, Girish
Kotwal, Sradha

Journal Title

Journal ISSN

Volume Title

Publisher

W B Saunders Co

Abstract

Rationale & Objective Hemodialysis catheter dysfunction is an important problem for patients with kidney failure. The optimal design of the tunneled catheter tip is unknown. This study evaluated the association of catheter tip design with the duration of catheter function. Study Design Observational cohort study using data from the nationwide REDUCCTION trial. Setting & Participants 4,722 adults who each received hemodialysis via 1 or more tunneled central venous catheters in 37 Australian nephrology services from December 2016 to March 2020. Exposure Design of tunneled hemodialysis catheter tip, classified as symmetrical, step, or split. Outcome Time to catheter dysfunction requiring removal due to inadequate dialysis blood flow assessed by the treating clinician. Analytical Approach Mixed, 3-level accelerated failure time model, assuming a log-normal survival distribution. Secular trends, the intervention, and baseline differences in service, patient, and catheter factors were included in the adjusted model. In a sensitivity analysis, survival times and proportional hazards were compared among participants’ first tunneled catheters. Results Among the study group, 355 of 3,871 (9.2%), 262 of 1,888 (13.9%), and 38 of 455 (8.4%) tunneled catheters with symmetrical, step, and split tip designs, respectively, required removal due to dysfunction. Step tip catheters required removal for dysfunction at a rate 53% faster than symmetrical tip catheters (adjusted time ratio, 0.47 [95% CI, 0.33-0.67) and 76% faster than split tip catheters (adjusted time ratio, 0.24 [95% CI, 0.11-0.51) in the adjusted accelerated failure time models. Only symmetrical tip catheters had performance superior to step tip catheters in unadjusted and sensitivity analyses. Split tip catheters were infrequently used and had risks of dysfunction similar to symmetrical tip catheters. The cumulative incidence of other complications requiring catheter removal, routine removal, and death before removal were similar across the 3 tip designs. Limitations Tip design was not randomized. Conclusions Symmetrical and split tip catheters had a lower risk of catheter dysfunction requiring removal than step tip catheters.Funding Grants from government (Queensland Health, Safer Care Victoria, Medical Research Future Fund, National Health and Medical Research Council, Australia), academic (Monash University), and not-for-profit (ANZDATA Registry, Kidney Health Australia) sources. Trial Registration Registered at ANZCTR with study number ACTRN12616000830493. Plain-Language Summary Central venous catheters are widely used to facilitate vascular access for life-sustaining hemodialysis treatments but often fail due to blood clots or other mechanical problems that impede blood flow. A range of adaptations to the design of tunneled hemodialysis catheters have been developed, but it is unclear which designs have the greatest longevity. We analyzed data from an Australian nationwide cohort of patients who received hemodialysis via a tunneled catheter and found that catheters with a step tip design failed more quickly than those with a symmetrical tip. Split tip catheters performed well but were infrequently used and require further study. Use of symmetrical rather than step tip hemodialysis catheters may reduce mechanical failures and unnecessary procedures for patients.

Description

Keywords

Citation

Source

American Journal of Kidney Diseases

Book Title

Entity type

Access Statement

License Rights

Restricted until

2099-12-31