Specific Anion effects on glass electrode pH measurements of Buffer solutions: Bulk and surface phenomena
Loading...
Date
Authors
Salis, Andrea
Pinna, M Cristina
Bilanicova, Dagmar
Monduzzi, Maura
Lo Nostro, Pierandrea
Ninham, Barry
Journal Title
Journal ISSN
Volume Title
Publisher
American Chemical Society
Abstract
The effect of electrolytes on pH measurements via glass electrodes is explored with solutions buffered at pH 7 (phosphate and cacodylate). Salt and buffer concentrations are varied. Direct and reverse Hofmeister effects are observed. The phenomena are significant for salt concentrations above 0.1 M and for buffer concentrations below 20 mM. Changes in measured pH show up most strongly with anions. They can be related to the usual physicochemical parameters (anion molar volumes, molar refractivity, and surface tensions) that are characteristic of Hofmeister series. They correlate strongly with anionic excess polarizabilities; this suggests the involvement of non-electrostatic, or dispersion, forces acting on ions. These forces contribute to ionic adsorption at the glass electrode surface, and to the liquid junction potential.
Description
Citation
Collections
Source
Journal of Physical Chemistry B
Type
Book Title
Entity type
Access Statement
License Rights
Restricted until
2037-12-31