Absolute calibration of Doppler coherence imaging velocity images
Loading...
Date
Authors
Samuell, Cameron
Allen, S L
Meyer, W
Howard, John
Journal Title
Journal ISSN
Volume Title
Publisher
IOP Publishing
Abstract
A new technique has been developed for absolutely calibrating a Doppler Coherence
Imaging Spectroscopy interferometer for measuring plasma ion and neutral velocities. An optical model of the interferometer is used to generate zero-velocity reference images for the plasma spectral line of interest from a calibration source some spectral distance away. Validation of
this technique using a tunable diode laser demonstrated an accuracy better than 0:2 km/s over an extrapolation range of 3.5 nm; a two order of magnitude improvement over linear approaches. While a well-characterized and very stable interferometer is required, this technique opens up the possibility of calibrated velocity measurements in difficult viewing geometries and for complex spectral line-shapes.