Calibrating a high-resolution wavefront corrector with a static focal-plane camera
Loading...
Date
Authors
Korkiakoski, Visa
Doelman, Neik
Codona, Johanan L
Kenworthy, M A
Otten, Gilles
Keller, Christoph U.
Journal Title
Journal ISSN
Volume Title
Publisher
Optical Society of America
Abstract
We present a method to calibrate a high-resolution wavefront (WF)-correcting device with a single, static camera, located in the focal-plane; no moving of any component is needed. The method is based on a localized diversity and differential optical transfer functions to compute both the phase and amplitude in the pupil plane located upstream of the last imaging optics. An experiment with a spatial light modulator shows that the calibration is sufficient to robustly operate a focal-plane WF sensing algorithm controlling a WF corrector with 40,000 degrees of freedom. We estimate that the locations of identical WF corrector elements are determined with a spatial resolution of 0.3% compared to the pupil diameter.
Description
Keywords
Citation
Collections
Source
Applied Optics
Type
Book Title
Entity type
Access Statement
Open Access