Evaluation of lead shielding for a gamma-spectroscopy system

Date

2008

Authors

Smith, Michael
Bignell, Lindsay Jordan
Alexiev, Dimitri
Mo, L
Harrison, J

Journal Title

Journal ISSN

Volume Title

Publisher

Elsevier

Abstract

Three types of lead shielding assemblies were investigated to establish which provides the lowest background as preliminary work for determining the efficiency of a High Purity Germanium (HPGe) detector. The background continuum and peaks were examined using gamma spectroscopy with a coaxial and a planar HPGe detector. The210Pb concentration in each lead sample taken from the assemblies was quantitatively determined. A French lead sample was found to have the lowest background in this relative comparison with a210Pb concentration of 29.1±1.3 Bq kg-1. Crown

Description

Keywords

Keywords: Concentration (process); Continuum mechanics; Gamma ray spectrometers; Particle detectors; Alpha spectrometry; High Purity Germanium (HPGe); Lead shield; Lead compounds 210Pb; Alpha spectrometry; Gamma spectrometry; Lead shield

Citation

Source

Nuclear Instruments and Methods in Physics Research: Section A

Type

Journal article

Book Title

Entity type

Access Statement

License Rights

Restricted until