Calibration and Standardisation of Synchrotron Radiation Circular Dichroism and Conventional Circular Dichroism Spectrophotometers

dc.contributor.authorMiles, Andrew J.en
dc.contributor.authorWien, Franken
dc.contributor.authorLees, Jonathan G.en
dc.contributor.authorRodger, A.en
dc.contributor.authorJanes, Robert W.en
dc.contributor.authorWallace, B. A.en
dc.date.accessioned2026-01-01T12:42:08Z
dc.date.available2026-01-01T12:42:08Z
dc.date.issued2003en
dc.description.abstractSynchrotron radiation circular dichroism (SRCD) is an emerging technique in structural biology with particular value in protein secondary structure analyses since it permits the collection of data down to much lower wavelengths than conventional circular dichroism (cCD) instruments. Reference database spectra collected on different SRCD instruments in the future as well as current reference datasets derived from cCD spectra must be compatible. Therefore there is a need for standardization of calibration methods to ensure quality control. In this study, magnitude and optical rotation measurements on four cCD and three SRCD instruments were compared at 192.5, 219, 290 and 490 nm. At high wavelengths, all gave comparable results, however, at the lower wavelengths, some variations were observable. The consequences of these differences on the spectrum, and the calculated secondary structure, of a representative protein (myoglobin) are demonstrated. A method is proposed for standardising spectra obtained on any CD instrument, conventional or synchrotron-based, with respect to existing and future databases.en
dc.description.statusPeer-revieweden
dc.format.extent9en
dc.identifier.issn0712-4813en
dc.identifier.otherORCID:/0000-0002-7111-3024/work/162949309en
dc.identifier.scopus0348147599en
dc.identifier.urihttps://hdl.handle.net/1885/733800398
dc.language.isoenen
dc.sourceSpectroscopyen
dc.titleCalibration and Standardisation of Synchrotron Radiation Circular Dichroism and Conventional Circular Dichroism Spectrophotometersen
dc.typeJournal articleen
dspace.entity.typePublicationen
local.bibliographicCitation.lastpage661en
local.bibliographicCitation.startpage653en
local.contributor.affiliationMiles, Andrew J.; Birkbeck University of Londonen
local.contributor.affiliationWien, Frank; Birkbeck University of Londonen
local.contributor.affiliationLees, Jonathan G.; Birkbeck University of Londonen
local.contributor.affiliationRodger, A.; University of Warwicken
local.contributor.affiliationJanes, Robert W.; Queen Mary University of Londonen
local.contributor.affiliationWallace, B. A.; Birkbeck University of Londonen
local.identifier.citationvolume17en
local.identifier.doi10.1155/2003/379137en
local.identifier.pured8ee585f-932f-4d60-8b30-cf0dae02ce5een
local.identifier.urlhttps://www.scopus.com/pages/publications/0348147599en