Continuously pumped rotating millikelvin cryostat

Loading...
Thumbnail Image

Date

Authors

Close, J. D.
Zieve, R. J.
Davis, J. C.
Packard, R. E.

Journal Title

Journal ISSN

Volume Title

Publisher

Access Statement

Research Projects

Organizational Units

Journal Issue

Abstract

We report the construction of a rotating nuclear demagnetization cryostat which can be continuously operated. The refrigerator employs a quadruple concentric rotating vacuum seal. The minimum 3He temperature of this cryostat in the stationary state is 165 μK. In the rotating state there is a relatively large heat leak which depends both on the rotation rate and on the field in the main demagnetization magnet.