Design and application of a high-precision, broad spectral range CCD-based absorption spectrometer with millisecond time resolution

Loading...
Thumbnail Image

Date

Authors

Steffen, Ronald
Jackman, Keith
Krausz, Elmars

Journal Title

Journal ISSN

Volume Title

Publisher

Institute of Physics Publishing

Abstract

We report the design, construction and performance of a CCD-based spectrometer system developed to study electronic absorbance spectral changes associated with electron transfer processes in biological systems. The instrument was designed to operate effectively over the 180-1100 nm wavelength range. The temperature of the sample can be controlled between 5 K and above 300 K. When operating between 500 nm and 1000 nm the resolution is ∼1.7-2.4 nm and stability better than 0.001 nm, whilst maintaining close to shot-noise-limited noise performance and minimal actinic fluence. Fully reproducible spectra can be taken in a single, gated exposure as short as 100 νs. Illustrative data taken of photosystem II core complexes are presented to demonstrate the spectrometer's ability to monitor complex multi-component kinetics with precision and speed. Laser-flash-induced spectral features, previously requiring laborious point-by-point accumulation and considerable signal averaging, can now be easily seen in a single measurement.

Description

Citation

Source

Measurement Science and Technology

Book Title

Entity type

Access Statement

License Rights

Restricted until

2037-12-31