Comparison of enzyme-linked immunosorbent assay and rapid chemiluminescent analyser in the detection of myeloperoxidase and proteinase 3 autoantibodies
Loading...
Date
Authors
Pucar, Phillippa A
Hawkins, Carolyn
Randall, Katrina
Li, Candice
McNaughton, Euan
Cook, Matthew
Journal Title
Journal ISSN
Volume Title
Publisher
Taylor & Francis Group
Abstract
Antibodies to myeloperoxidase (MPO) and proteinase 3 (PR3) are vital in the diagnosis and management of ANCA-associated vasculitis. A chemiluminescent immunoassay (CLIA; Quanta Flash) provides MPO and PR3 antibody results in 30 minutes, which is much faster than enzyme-linked immunosorbent assay (ELISA). We compared the performance of ELISA (Orgentec) and CLIA (Quanta Flash) for MPO and PR3 antibody quantitation on 303 samples, comprising 196 consecutive samples received in a single diagnostic laboratory over a 3 month period, and 107 samples collected from 42 known vasculitis patients over a 40 month period. We observed a correlation between both methods using spearman correlation coefficients (MPO, rs = 0.63, p < 0.01; PR3, rs = 0.69, p < 0.01). There was agreement between both methods in determining a positive or negative result. In the vasculitis cohort, CLIA performed well at clinically important stages of disease; diagnosis (eight samples all positive by both assays) and disease relapse (correlation for both MPO and PR3 antibody quantitation rs = 0.84, p = 0.03 and rs = 0.78, p < 0.01, respectively). Three samples were discordant at clinical relapse, testing positive by CLIA, including one high positive associated with relapse requiring a change in treatment. In summary, CLIA appears to be at least as accurate as ELISA for measurement of MPO and PR3 antibodies
Description
Keywords
Citation
Collections
Source
Pathology