Skip navigation
Skip navigation

Capacitance Dilatometry Attachment for a Surface-Force Apparatus

Stewart, Andrew

Description

An application of capacitance dilatometry to a surface-force apparatus is described. The distance between the surfaces is obtained from the capacitance of a cylindrical parallel-plate capacitor measured with a variable-ratio transformer bridge; the surface separation is made to vary with a magnetic force transducer. The method permits measurements of forces between opaque surfaces over periods of time ranging from tens of milliseconds to, in principle, an indefinite period. A new method of...[Show more]

CollectionsANU Research Publications
Date published: 2000
Type: Journal article
URI: http://hdl.handle.net/1885/89804
Source: Measurement Science and Technology
DOI: 10.1088/0957-0233/11/3/319

Download

There are no files associated with this item.


Items in Open Research are protected by copyright, with all rights reserved, unless otherwise indicated.

Updated:  12 November 2018/ Responsible Officer:  University Librarian/ Page Contact:  Library Systems & Web Coordinator