The comparative analytical performance of four troponin I assays at low concentration

Koerbin, Gus; Tate, J R; Potter, Julia; Hickman, Peter

Description

Background: Low troponin concentrations have been shown to be informative in the prognosis of acute coronary syndrome. We have investigated the analytical performance of four commonly used cardiac troponin I methods at concentrations approaching their analytical limit of detection. Method: We assayed 167 patient samples within 24 h of collection using the Beckman Coulter AccuTnl, Dade Behring Dimension, Abbott AxSYM and Bayer Centaur methods and compared their relative analytical performance....[Show more]