Skip navigation
Skip navigation

Origins of the Residual Pulse Height Deficit in Propane-Filled Gas Ionization Detectors

Dall (previously Weijers), Tessica; Timmers, Heiko; Elliman, Robert

Description

This work investigates the origins of the residual pulse height deficit in gas ionization detectors. It is motivated by the recent observation that the species dependence of gas detector response cannot be accounted for solely by considering the energy loss of the ions in the detector window and non-ionizing energy loss processes in the detector gas. It was found that the residual pulse height deficit is approximately proportional to the square of the ionization density. However, only a weak...[Show more]

CollectionsANU Research Publications
Date published: 2005
Type: Journal article
URI: http://hdl.handle.net/1885/73711
Source: Nuclear Instruments and Methods in Physics Research: Section A