Quality assurance in accelerator mass spectrometry: Results from an international round-robin exercise for 10 Be

Merchel, S.; Bremser, W.; Akhmadaliev, S; Arnold, M.; Aumaitre, G.; Bourles, D.L.; Braucher, R; Caffee, M.; Christl, M; Fifield, L Keith; Finkel, Robert C; Freeman, S P H T; Tims, Stephen; Wallner, Anton; Freeman, S; Martschini, M.; Wilcken, Klaus M.; Xu, S


The first international round-robin exercise for the measurement of the long-lived radionuclide 10Be has been conducted. Ten participating accelerator mass spectrometry (AMS) facilities have each measured three samples at the 10 -12 to 10 -1410Be/ 9Be level. All results have been made traceable to the NIST SRM 4325 standard to avoid additional discrepancies that arise when different facilities use different calibration materials. Hence, the data concentrates on pure measurement distinctions....[Show more]