Skip navigation
Skip navigation

Reliable automatic alignment of tomographic projection data by passive auto-focus

Kingston, Andrew; Sakellariou, Arthur; Varslot, Trond; Myers, Glenn; Sheppard, Adrian

Description

Purpose: The authors present a robust algorithm that removes the blurring and double-edge artifacts in high-resolution computed tomography (CT) images that are caused by misaligned scanner components. This alleviates the time-consuming process of physically aligning hardware, which is of particular benefit if components are moved or swapped frequently. Methods: The proposed method uses the experimental data itself for calibration. A parameterized model of the scanner geometry is constructed and...[Show more]

CollectionsANU Research Publications
Date published: 2011
Type: Journal article
URI: http://hdl.handle.net/1885/59987
Source: Medical Physics
DOI: 10.1118/1.3609096

Download

File Description SizeFormat Image
01_Kingston_Reliable_automatic_alignment_2011.pdf3.55 MBAdobe PDF    Request a copy


Items in Open Research are protected by copyright, with all rights reserved, unless otherwise indicated.