In situ measurement of hafnium isotopes in rutile by LA-MC-ICPMS: Protocol and applications

Ewing, Tanya; Rubatto, Daniela; Eggins, Stephen; Hermann, Joerg

Description

This work presents a rigorous assessment of the accuracy and precision with which176Hf/177Hf can be measured in rutile by laser ablation (LA) MC-ICPMS. For rutile with ≥40ppm Hf, we demonstrate that176Hf/177Hf can be measured accurately and reproducibly

CollectionsANU Research Publications
Date published: 2011
Type: Journal article
URI: http://hdl.handle.net/1885/27019
Source: Chemical Geology
DOI: 10.1016/j.chemgeo.2010.11.029