Skip navigation
Skip navigation

Comparison of enzyme-linked immunosorbent assay and rapid chemiluminescent analyser in the detection of myeloperoxidase and proteinase 3 autoantibodies

Pucar, Phillippa A; Hawkins, Carolyn; Randall, Katrina; Li, Candice; McNaughton, Euan; Cook, Matthew

Description

Antibodies to myeloperoxidase (MPO) and proteinase 3 (PR3) are vital in the diagnosis and management of ANCA-associated vasculitis. A chemiluminescent immunoassay (CLIA; Quanta Flash) provides MPO and PR3 antibody results in 30 minutes, which is much faster than enzyme-linked immunosorbent assay (ELISA). We compared the performance of ELISA (Orgentec) and CLIA (Quanta Flash) for MPO and PR3 antibody quantitation on 303 samples, comprising 196 consecutive samples received in a single diagnostic...[Show more]

CollectionsANU Research Publications
Date published: 2017
Type: Journal article
URI: http://hdl.handle.net/1885/217915
Source: Pathology
DOI: 10.1016/j.pathol.2017.02.006