Calibration of high-temperature furnace assemblies for experiments between 200 and 600 MPa with end-loaded piston-cylinder apparatuses
-
Altmetric Citations
Farias Vlach, Silvio Roberto; Salazar-Naranjo, Andrés Fabián; Torres-Corredor, Johan Santiago; de Carvalho, Paulo Rafael; Mallmann, Guilherme
Description
We present pressure calibration results for piston-cylinder –” and 1” high temperature furnace assemblies in the 200-600 MPa range aiming to expand its applicability to simulate upper crustal conditions. The furnace assemblies were made up by crushable MgO, graphite heater, pyrex glass spacer, and NaCl sleeve. Twenty-one experiments were carried out over a range of pressures (200-600 MPa) and temperatures (870-945°C) to determine the liquidus curve of NaCl. The obtained curve slope fits that...[Show more]
Collections | ANU Research Publications |
---|---|
Date published: | 2019-04-15 |
Type: | Journal article |
URI: | http://hdl.handle.net/1885/196756 |
Source: | Brazilian Journal of Geology |
DOI: | 10.1590/2317-488920192018090 |
Access Rights: | Open Access |
Download
File | Description | Size | Format | Image |
---|---|---|---|---|
01_Farias+Vlach_Calibration_of_2019.pdf | 918.71 kB | Adobe PDF |
This item is licensed under a Creative Commons License