Skip navigation
Skip navigation

Three view electronically scanned interferometer for plasma electron density measurements on the H-1 heliac

Oliver, David; Howard, John; Kumar, Santhosh T. A.; Pretty, D. G.; Blackwell, B. D.

Description

We report the development of a three view electronically scanned millimeter-wave interferometer for plasma electron density profile measurement on the H-1 heliac. The system utilizes an electronically tunable backward-wave oscillator whose output is incident on a fixed blazed diffraction grating such that sweeping the source frequency effects a spatial scan of the plasma cross section. Two diagonal views essentially span most of the plasma cross section, while the horizontal arm views the lower...[Show more]

CollectionsANU Research Publications
Date published: 2006-09-27
Type: Journal article
URI: http://hdl.handle.net/1885/16368
Source: Review of Scientific Instruments
DOI: 10.1063/1.2218839

Download

File Description SizeFormat Image
01_Oliver_Three_view_electronically_2006.pdfPublished Version629.63 kBAdobe PDFThumbnail


Items in Open Research are protected by copyright, with all rights reserved, unless otherwise indicated.