3D Silicon Microdosimetry and RBE Study Using C-12 Ion of Different Energies

Tran, Linh T.; Chartier, Lachlan; Bolst, David; Prokopovich, Dale A.; Guatelli, Susanna; Nancarrow, Mitchell; Reinhard, Mark I.; Petasecca, Marco; Lerch, Michael L.F.; Hinde, David; Pereverlaylo, Vladimir L.; Dasgupta, Mahananda; Matsufuji, Naruhiro; Stuchbery, Andrew; Jackson, Michael; Rozenfeld, Anatoly Borisovich

Description

This paper presents a new version of the 3D mesa Bridge microdosimeter comprised of an array of 4248 silicon cells fabricated on 10 μm thick n-type silicon-on-insulator substrate. This microdosimeter has been designed to overcome limitations existing in previous generation silicon microdosimeters and it provides well-defined sensitive volumes and high spatial resolution. The charge collection characteristics of the new 3D mesa microdosimeter were investigated using the ANSTO heavy ion...[Show more]

CollectionsANU Research Publications
Date published: 2015
Type: Journal article
URI: http://hdl.handle.net/1885/149074
Source: IEEE Transactions on Nuclear Science
DOI: 10.1109/TNS.2015.2487266

Download

File Description SizeFormat Image
01_Tran_3D_Silicon_Microdosimetry_and_2015.pdf1.86 MBAdobe PDF    Request a copy


Items in Open Research are protected by copyright, with all rights reserved, unless otherwise indicated.

Updated:  22 January 2019/ Responsible Officer:  University Librarian/ Page Contact:  Library Systems & Web Coordinator